Vertical sprinkler
SAN-7103AS

The spray switch ball valve of the vertical sprinkler can be quickly started within 1 second, providing water source quickly in emergency situations and saving valuable treatment time for users

Made of 316 stainless steel material, the inner wall of the pipe is smooth mirror surface, free of grease and rust

Product Description

Vertical sprinkler is an indispensable emergency rescue device in laboratory working environment. In emergency situations, when exposed body parts of workers come into contact with toxic, harmful, or corrosive chemicals, a vertical sprayer can be used for full body flushing to provide effective initial rescue and protect the health and safety of workers.

The production of SAN series vertical sprayers produced by Taixiong strictly follows the national standard GB 38144-2025 "Eye and Face Protection - Emergency Sprinkler and Eye Washing Equipment" ANSI/ISEA Z358.1、 And it has been certified by the international authoritative TUV, proving its compliance with relevant safety and performance standards.

认证.png

Product Parameters

Subject
316 Stainless Steel
Coating
Passivation treatment
Sign
PVC
Spray plate
316 Stainless Steel
Net weight
10.80kg
Pipe diameter
45mm

Product Drawing

Precautions

Layout
The equipment should be set within a 10 second walking distance from potential hazardous areas, and it is recommended to maintain a straight-line distance between 10-15 meters. Ensure that the path from the hazardous area to the equipment is unobstructed and straight, in order to quickly reach and use it in emergency situations
Spray range
There are no obstacles within 410mm of the center of the spray range
Utilities
The dynamic water pressure should be between 0.2Mpa-0.4Mpa
Rinse requirements
Spray flow rate ≥ 76 L/min, continuous flushing should be carried out for at least 15 minutes
Wastewater treatment
Please dispose of the wastewater promptly after use to avoid the formation of puddles or freezing at low temperatures, in order to prevent slipping after use
Training requirements
Before putting the vertical sprinkler into use, please train the practitioners on the correct use of the device
Water quality temperature
Use pure water at room temperature or water that meets hygiene standards, with a suitable temperature range of 16 ℃ -38 ℃. The water temperature should not be too high to avoid accelerating the reaction of chemical substances and causing accidents